3.8 Error analysis of X-RAY control signal path: up to unit hardware revision AG
H321
is triggered at start of exposure.
3.8.2
Error messages
without installed remote
control
E3 13 40 + E3 41 20
occur in combination after the unit
is switched ON.
Door contact
open?
• Close door contact
• Start exposure again
Is the error message displayed
again?
yes
yes
no
• Check door contact, repair if necessary
• Replace cable L17, (see section 6.21)
yes
yes
no
Was the release but-
ton actuated during
switch ON?
Switch the unit OFF
and then ON again.
Make sure that the
release button is not
actuated during
switch-ON.
Do the error messages
recur?
yes
Short circuit in coiled cable of release button:
• Replace release button (unit hardware version
<
BA)
• Replace board DX41 (see section 6.20)
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...