![background image](http://html1.mh-extra.com/html/sirona/orthophos-xg-3-ds/orthophos-xg-3-ds_service-manual_1285016141.webp)
59 38 399 D3352
D3352.076.01.13.02
07.2008
3 – 17
Tab 3
3.7 Checking the cables
3.7
3.7
Checking the cables
NOTE
i
You can use a standard Cat5 cable as a test cable for L8*, L10, L12, L40
and L37. Caution! This cable must not be permanently installed.
NOTE
i
Most cables have the same plug at both ends and are connected 1:1
(see also section "1.12 Cabling overview').
* up to unit hardware version AG
Is the cable plug connection OK?
For shielded cables:
Is the cable shield connection OK?
yes
If a substitute cable is available,
temporarily connect it and
check unit functioning.
• Is the unit functioning properly?
yes
Replace cable, see section 6.21.
Clean plug; in case of a defect,
fix plug, replace cable if necessary,
see section 6.21.
no
Repair shield,
replace cable if necessary,
see section 6.21.
no
Locate module causing fault and
replace component,
see section 6.
no
yes
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...