6.9 Replacing/adjusting the FH light localizer (Ceph)
6.9
Adjusting the Ceph light localizer (FH)
z
Switch the unit
ON
.
z
Switch the light localizers
ON
.
z
Align the Ceph (FH) laser module so that FH light beam
C
runs horizontal
between the ear plugs (tolerance: ± 1.5 mm).
z
Tighten screw
A
securely.
NOTE
i
Be careful not to turn the laser module when you tighten the screws.
z
Check the light localizer again and perform a correction if necessary.
Attaching the covers
z
Reattach the cover of the secondary diaphragm.
z
Reattach the cover to the cephalometer.
ATTENTION: What to do after replacement?
– No further action is required.
60
90
120
70
80
100 110
C
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...