59 38 399 D3352
5 – 50
D3352.076.01.13.02
07.2008
5.8 Service routine S008
Tabs 5
S008.3: Input/confirmation of unit serial number
5.8.2
S008: Test step 3
Selecting test step 3
1.
Select test step 3 in selection field 2 with the
arrow keys
and confirm your
selection by pressing the
Service key
.
This test step is required...
z
if a software update is performed...
for XG 5 / 3 / 3 PPE: from V02.18.1 to a higher version
for XG
Plus
:
from V02.19 to a higher
version.
The value X is then displayed in selection field 1 of service routine
S008.3 (and with the arrow keys for the XG
Plus
). When opened, the
routine is initially in the editing mode for entry of the unit serial number
(case A).
z
if modules have been replaced. In this case, the backup copy of the unit
serial number still stored in the unit must be reconfirmed (case B).
If this service routine is called up with a valid unit serial number, the possibility
of entering or confirming the serial number will be deactivated. The valid serial
number is displayed in selection field 1.
1
2
3
+
-
S008
3
n
(
1
2
3
S008_______3____
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE:
Multipad
+
-
(
n
(
n
(
o
(
o
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...