4.3 Adjusting the panoramic X-ray unit
Tabs 4
Making the unit ready for exposure
NOTE
i
The default values for S1, S2 and S3 were automatically determined by
SIDEXIS based on the exposure and entered in the text boxes of the menu.
For manual adjustment, the values displayed at this position in the text boxes
of the menu can be overwritten (see page 4-21).
First continue with the automatic adjustment. Manual determination of the ad-
justment values is required only if you fail to reach your goal via automatic ad-
justment (see page 4-20).
5.
Make SIDEXIS XG ready for exposure:
Click
IMAGE
ACQUISITION
The
PAN
-
SENSOR
ADJUSTMENT
dialog box appears in SIDEXIS.
The dialog box suggests a value
Lx
for the mechanical adjustment of the
sensor.
If the suggested value is greater than ±0.5mm, then perform a mechani-
cal adjustment of the sensor.
Moving the sensor
6.
Adjust the sensor:
NOTE
i
Positive sign
= moves the sensor to the right
Negative sign
= moves the sensor to the left
– Mark the position of the sensor.
– Loosen the three screws
A
(approx. 2 turns, do not unscrew fully)
– Move the sensor to the left or right by the displayed value
Lx
.
– Tighten screws
A
again.
A
A
A
+
-
System version 1: Fixed sensor holder
A
A
A
Hardware version 2: motor-driven sensor holder
+
-
(XG
Plus
, TSA only)
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...