2.5 List of error messages
Tab 2
E5
13
13
Malfunction of actuator 2
during operation
z
Check the actuator mechanisms manually for
smooth and easy running.
In case of stiffness...
z
manually reset the actuators to the zero position
and run a test cycle via P1.
If the error occurs again...
z
Check light barrier V1_2 (X803) with service
routine S015.5,
replace light barrier if necessary
z
Check electric drive and jolt-free running with
service routine S015.5 (free travel),
z
replace the actuators and/or spindles and
replace board DX1 if necessary
E4
13
14
Actuator 2;
position counter error
z
Check the actuator mechanisms manually for
smooth and easy running.
In case of stiffness...
z
manually reset the actuators to the zero position
and run a test cycle via P1.
If the error occurs again...
z
Check light barrier V1_2 (X803) with service
routine S015.5,
replace light barrier if necessary
z
Check electric drive and jolt-free running with
service routine S015.5 (free travel),
z
replace the actuators and/or spindles and
replace board DX1 if necessary
E6
13
15
Actuator 2 is not ready for operation
This error is a sequential fault.
z
Unit restart:
Switch the unit OFF. Wait 30 sec.
Switch unit ON.
z
Repeat procedure and observe causal error
messages.
E6
13
17
Error when activating actuator 2
z
Unit restart:
Switch the unit OFF. Wait 30 sec.
Switch unit ON and check functioning.
z
Please contact the Sirona Customer Service
Center (CSC) to find out whether a bugfix by
means of a software update is possible and per-
form such an update if necessary.
E4
13
21
Ring motor has not reached pan home
position
z
Check the actuator mechanisms manually for
smooth and easy running,
reset actuators to zero position of necessary.
z
Check the ring drive mechanism manually for
smooth and easy running,
replace the ring motor or mechanism if neces-
sary
z
Check light barrier V1_3 (X804) with service
routine S014.4,
Error code
Description
Actions required
see page
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...