59 38 399 D3352
4 – 4
D3352.076.01.13.02
07.2008
4.1 Important information concerning adjustment
Tabs 4
Default values in the Diaphragm/system adjustment menu in
SIDEXIS XG
During the adjustment, the default adjustment values are displayed in the text
boxes of the Diaphragm/system adjustment menu.
First perform the adjustment with these default values. If you do not attain the
desired result via this automatic adjustment, you should determine the adjust-
ment values manually by measuring the exposure with the SIDEXIS measur-
ing ruler and then overwrite the default values in the menu.
This procedure is described in the following sections.
Return ring travel with unit SW version V02.27 and higher
Starting with unit version V02.27, return ring travel is automatically performed
during service operation (during adjustment). Actuation of the R key before or
after each adjustment exposure (as described in the following chapters) can
thus be omitted.
NOTE
i
It is not possible to select a new exposure during the return ring travel
(incl. resetting of the actuators)!
If a selection is made during this period, the error message "It is not possible to
set the adjustment values at this time" or “No exposure can be taken" appears.
Acknowledge the error message and wait until the return ring travel has been
completed.
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...