4.3 Adjusting the panoramic X-ray unit
4.
3
The fixed diaphragm must be dismantled in order to perform the sensor
adjustment of the ORTHOPHOS XG 3/3 PPE X-ray unit. To do this, proceed
as follows:
1.
Remove the covers:
– Tube assembly, front
– Tube assembly, rear
2.
Loosen the two screws
(B)
(approx. 2-3 turns) and push the fixed
diaphragm upward.
z
Assemble by following the same procedure in reverse order.