59 38 399 D3352
6 – 84
D3352.076.01.13.02
07.2008
6.20 Replacing circuit boards
Tabs 6
11.
Open the Service menu (see page 5-10).
12.
Start service routine
S008.3
, check the unit serial number and confirm it if
it is correct (see page 5-50).
NOTE
i
The unit serial number is located on the rating plate of the unit.
CAUTION
In case of a wrong serial number, cancel the update and contact the Sirona
Customer Service Center!
13.
Switch the unit OFF.
Wait approximately 1 minute. Then switch the unit ON again.
14.
Install the current SIDEXIS software version
(V01.53 or higher).
NOTE
i
If the current SIDEXIS version is a patch version, the previous officially
released main version of SIDEXIS has to be installed before the current
version can be installed.
15.
Perform a software update to Version V02.27 or higher (automatic up-
date) (see also section 1.9.5).
16.
Perform the complete system adjustment (see chapter 4):
– PAN adjustment
– For TSA units: TSA adjustment
– CEPH adjustment if cephalometer is installed
17.
Perform an acceptance test (applies to Germany only);
without consulting an expert
18.
Open the "Extended Details“ via SIXABCON.
This generates an XML file (with the system parameters) which is saved
under the network name of the system in the PDATA/P2K_Config folder
(see also section 1.10 on page 1-32).
z
The process is completed.
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...