4.3 Adjusting the panoramic X-ray unit
Tabs 4
4.3.3
Adjusting the pan sensor
(XG
Plus
/ XG 5 / 3 / 3 PPE)
Menu: Diaphragm/system adjustment
‡
Pan - Sensor adjustment
Preparations
z
Plug the sensor into the sensor slot on the panoramic X-ray unit.
z
Insert the needle phantom in the bite block holder of the panoramic X-ray
unit (see page 4-9).
ORTHOPHOS XG 3 / 3 PPE:
z
The fixed diaphragm must be dismantled prior to the sensor adjustment
for the ORTHOPHOS XG 3 / 3 PPE system version (see page 4-10).
Opening
PAN
-
SENSOR
ADJUSTMENT
1.
Open the
PAN
-
SENSOR
ADJUSTMENT
submenu (see section 4.2).
NOTE
i
The menu provides a precision adjustment and a coarse adjustment
(precision adjustment is preset). Perform a precision adjustment first. In most
cases, previous coarse adjustment is not necessary.
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...