båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
4 – 5
Tab 4
4.2 Diaphragm/system adjustment menu
4.
2
4.2
Diaphragm/system adjustment menu
Menu: Diaphragm/system adjustment
The
DIAPHRAGM
/
SYSTEM
ADJUSTMENT
menu guides you through the
adjustment of the panoramic unit and the cephalometer. This service routine
is started from
SIDEXIS XG
:
EXTRAS
Æ
CONSTANCY
TEST
Æ
XCXP
Æ
SELECT
X
-
RAY
DEVICE
Æ
SERVICE
EXPOSURE
Æ
DIAPHRAGM
/
SYSTEM
ADJUSTMENT
NOTE
i
The
DIAPHRAGM
/
SYSTEM
ADJUSTMENT
menu is password-protected. As
password, enter the first four digits of the current system date (PC) in reverse
order.
Example: On 05/30/2004, the service password is
5003
NOTE
i
When you open the
DIAPHRAGM
/
SYSTEM
ADJUSTMENT
menu, the unit
switches from the user mode to the PC service mode logged by the PC.
ORTHOPHOS XG
Plus
/ Ceph:
This mode is indicated by the
PC service screen
displayed on the Easypad.
In the PC service mode,
the control options that are available on the Easypad
are determined by SIDEXIS and the currently selected service routine. Gen-
eral control of the unit by means of the Easypad (as in the user mode) is not
possible in this mode.
ORTHOPHOS XG 5 / Ceph bzw. XG 3 / 3 PPE:
This mode is indicated by the
SERVICE
reading on the Multipad.
In the PC service mode,
the Multipad control options are determined by
SIDEXIS and the currently selected service routine. General control of the unit
by means of the Multipad (as in the user mode) is not possible in this mode.
XG
Plus
XG 5 / 3 / 3 PPE
SERVICE
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...