5.13 Service routine S017
Tabs 5
S017.14: Enabling/disabling certain lines of the welcome screen
Quitting the service routine
To go on to the next test step, press the Service key
.
To return to the Service menu, press the
double arrow key
.
To quit the Service menu and return to the main menu, press the
double arrow key
.
5.13.12
S017: Test step 14
(XG
Plus
full version only)
z
Start test step 14 in the same way as described on page 5-78 or by
pressing the
Service key
in test step 13.
NOTE
i
If you go to test step 14 from test step 13 by pressing the Service key, the
security access confirmation will be skipped.
After the test step is selected, a welcome screen line ID appears in selection
field 1.
1
2
3
+
-
1
XG
Plus
:
Easypad touchscreen
R
14
17
+
-
0
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...