båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
6 – 17
Tab 6
6.3 Replacing the PAN actuators (M1_1/2)
6.3
6.3
Replacing the PAN actuators (M1_1/2)
Removing the covers
z
Remove the covers (see section 1.14):
– Top arm cover
Removing the defective actuator
1.
Loosen the actuator cable from the cable holders and pull it off of connec-
tors
X811
(AK1; M1_1)
and
X812
(AK2; M1_2) on board
DX1
.
2.
Loosen set screws
A
on the coupling and the two screws
B
on the actua-
tor holder and pull the actuator including the holder out toward the rear.
Reusing the motor holder (if required)
3.
Loosen the four screws
(C)
and remove the actuator holder from the
defective motor.
z
Place the new actuator in the holder and fasten it with the four screws
C
.
M1_1 = right
M1_1 = left
A
B
C
NOTE
i
Seal set screws
A
with Loctite 242 before
fastening them tight.
1.
3.
2.
DX1
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...