båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
5 – 69
Tab 5
5.11 Service routine S014
5.11
S014.2: Travel of rotational drive to the ceph home position
Quitting the service routine
XG
Plus
:
To return to the service routine selection menu, press the
Service key
or the
double arrow key
.
To quit the Service menu and return to the main menu, press the
double arrow key
.
For XG 5 / 3 / 3 PPE:
Pressing the
up arrow key
above selection field 3 returns you to the ser-
vice routine selection menu.
To quit the Service menu and return to the Main menu, press the
up arrow key
above selection field 3.
5.11.2
S014: Test step 2
This test step realizes "isolated" travel of the rotation motor to the ceph home
position (with the arm in the right or the left position, depending on the config-
uration). Actuators 1 and 2 remain in their current positions.
Travel to the ceph home position is also possible without configuring the sys-
tem version for the cephalometer mode.
Selecting test step 2
1.
Select test step 2 in selection field 2 with the
arrow keys
and confirm your
selection by pressing the
Service key
.
1
2
3
+
-
S014
2
n
(
1
2
3
S014_______2____
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE:
Multipad
+
-
(
n
(
n
(
o
(
o
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...