59 38 399 D3352
2 – 64
D3352.076.01.13.02
07.2008
2.6 List of available service routines
Tab 2
S017
Test step 18
(V02.28 or higher, for XG
Plus
only):
in preparation!
VORSICHT
This routine must not be selected, since it would
cause malfunctions of the unit!
Test step 22
(V02.28 or higher, for XG 3 PPE
only):
Configuration of PPE display
z
Configuration of display of remaining exposures
after power-on
Test step 23
(V02.28 or higher, for XG 3 PPE
only):
Configuration of warning level for PPE
display
z
Configuration of warning level for display of re-
maining exposures after an exposure
S018
Set travel height
Test step 2:
Limit maximum travel height
Test step 3:
Undo limit of maximum travel height
setting
z
if room height is lower than 2.27 m
(2.30 m with floor stand)
Test step 4:
Test of the height adjustment sensor
system
z
Check of the sensor system for the height ad-
justment (upper and lower limit switches, cor-
rection switch, pulse counter)
Test step 5 (V02.27 and higher):
Limit minimum travel height
Test step 6 (V02.27 and higher):
Undo limit of minimum travel height
setting
z
if the minimum travel height must be limited
S020
Service for temple support
Test step 1
(V02.22 or higher, XG
Plus
only):
Activate/deactivate switching thresh-
olds of the temple supports
z
in case of complaints regarding the allocation of
the temple support distances to the jaw widths
S021
Diaphragm service: motor-driven dia-
phragm
Test step 1:
Initialization of diaphragm axis
z
following replacement of diaphragm unit
Test step 3:
Function test of
TSA sensor adjustment
z
Important for TSA Ready units:
after replacing the sensor holder
S032
Test function for sensor in PAN slot
Test step 10 (from V02.27):
Test function for sensor in PAN slot
z
for troubleshooting HSI problems
(E6 14 10, E6 14 11, E6 14 12)
S033
Test of CEPH image path without
SIDEXIS
Test step 10 (V02.27 or higher,
not for XG 3 / 3 PPE):
Test of CEPH image path without
SIDEXIS
z
for troubleshooting HSI problems
(E6 14 10, E6 14 11, E6 14 12)
Routine
Description
Required...
see page
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...