båÖäáëÜ
59 38 399 D3352
D3352.076.01.13.02
07.2008
2 – 39
Tab 2
2.5 List of error messages
2.5
E6
42
20
Contact to DX11 interrupted during
operation
z
Check the signal path from board DX1 to board
DX42,
replace module, if necessary
z
Check connection of remote control,
see installation instructions
z
Checking the CAN bus
z
Check cable L17,
replace if necessary
z
Check board DX42,
replace if necessary
z
Check board DX41,
replace if necessary
HINWEIS
i
If the error cannot be eliminated immediately, the
unit can be temporarily reconfigured and operated
with a release button located directly on the unit (see
installation instructions)
E7
42
21
No CAN bus connection.
DX11 does not start.
Note:
Occurs after power-on in the start
screen.
z
Check configuration (with or without DX41) via
service routine S017.9,
configure correctly if necessary
z
Check the signal path from board DX1 to board
DX42,
replace module if necessary
z
Checking the CAN bus
z
Check remote control via service routine 17.6,
configure if necessary
z
Start the detail query via Sixabcon
z
If DX11 responds...
z
Check the signal path to DX42,
repair or replace cable/connector if necessary
z
Replace DX1
If DX11 does not respond...
z
Replace DX11 if this error persists
E3
42
30
R key actuated during power-on
z
Unit restart:
Switch the unit OFF. Wait for 30 sec.
Switch unit ON, making sure that the remote
control is not actuated during boot-up.
If the error occurs again...
z
Replace remote control,
see installation instructions
Error code
Description
Actions required
see page
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...