Configuration
R&S
®
FSW-K7
67
User Manual 1173.9240.02 ─ 17
Comment
An optional comment that describes the conversion loss table. The comment can be
freely defined by the user.
Remote command:
Band
The waveguide or user-defined band for which the table is to be applied. This setting is
checked against the current mixer setting before the table can be assigned to the
range.
For a definition of the frequency range for the pre-defined bands, see
Remote command:
Harmonic Order
The harmonic order of the range for which the table is to be applied. This setting is
checked against the current mixer setting before the table can be assigned to the
range.
Remote command:
Bias
The bias current which is required to set the mixer to its optimum operating point. It
corresponds to the short-circuit current. The bias current can range from -10 mA to 10
mA. The actual bias current is lower because of the forward voltage of the mixer
diode(s).
Tip:
You can also define the bias interactively while a preview of the trace with the
changed setting is displayed, see
Remote command:
Mixer Name
Specifies the name of the external mixer for which the table is to be applied. This set-
ting is checked against the current mixer setting before the table can be assigned to
the range.
Remote command:
Mixer S/N
Specifies the serial number of the external mixer for which the table is to be applied.
This setting is checked against the current mixer setting before the table can be
assigned to the range.
Remote command:
Input and Frontend Settings