Configuration
R&S
®
FSW-K7
101
User Manual 1173.9240.02 ─ 17
Pulse Length ← Output Type ← Trigger 2/3
Defines the length of the pulse sent as a trigger to the output connector.
Remote command:
Send Trigger ← Output Type ← Trigger 2/3
Sends a user-defined trigger to the output connector immediately. Note that the trigger
pulse level is always opposite to the constant signal level defined by the output "Level"
setting, e.g. for "Level = High", a constant high signal is output to the connector until
the "Send Trigger" button is selected. Then, a low pulse is sent.
Which pulse level will be sent is indicated by a graphic on the button.
Remote command:
5.6 Data Acquisition
How data is to be acquired and then demodulated is configured in the "Data Acquisi-
tion" dialog box.
MSRA/MSRT operating mode
In MSRA/MSRT operating mode, only the MSRA/MSRT Master channel actually cap-
tures data from the input signal. The data acquisition settings for the Analog Demodu-
lation application in MSRA/MSRT mode define the analysis interval.
For details on the MSRA operating mode see the R&S
FSW MSRA User Manual. For
details on the MSRT operating mode see the R&S
FSW Realtime Spectrum Applica-
tion and MSRT Operating Mode User Manual.
............................................................................................... 101
5.6.1 Bandwidth Settings
The bandwidth settings define which parts of the input signal are acquired and then
demodulated. They are configured via the BW key or in the "Bandwidth" tab of the
"Data Acquisition" dialog box.
► To display this dialog box, do one of the following:
● Select the "Data Acquisition" button in the Analog Demodulation "Overview"
● Select the BW key and then the "Bandwidth Config" softkey.
Data Acquisition