![Rohde & Schwarz FSW Series User Manual Download Page 117](http://html.mh-extra.com/html/rohde-and-schwarz/fsw-series/fsw-series_user-manual_1477844117.webp)
Configuration
R&S
®
FSW-K7
117
User Manual 1173.9240.02 ─ 17
5.8.4 Scaling
The scaling parameters define the range of the demodulated data to be displayed.
They are configured in the "Scaling" tab of the "Demodulation Settings" dialog box.
► To display this dialog box, do one of the following:
● Select the "Demod Settings" button in the Analog Demodulation "Overview"
and select the "Scaling" tab.
● Select the MEAS CONFIG key and then the "Scale Config" softkey.
Depending on the evaluation (AF or RF display), the settings vary.
5.8.4.1
AF Evaluation
These settings are only available for AF evaluations.
Dev per Division/ Db per Division
................................................................................117
............................................................................................118
Dev per Division/ Db per Division
Defines the modulation depth or the phase deviation or frequency deviation per divi-
sion (logarithmic: 0.1 to 20 dB):
AM display:
0.0001 % to 1000 %
FM display:
1 Hz/div to 100 MHz/div
PM display:
0.0001 rad/div to 1000 rad/div
Demodulation