Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
222
User Manual 1173.9240.02 ─ 17
Parameters:
<Offset>
<numeric value>, specified in Hz, kHz, MHz or GHz, rounded to
the nearest Hz
*RST:
0 Hz
Example:
SOUR:EXT:FREQ:OFFS 10HZ
Sets an offset of the generator output frequency compared to
the analyzer frequency of 10 Hz.
Manual operation:
See
"(Automatic) Source Frequency (Numerator/Denominator/
SOURce:EXTernal:POWer[:LEVel]
<Level>
This command sets the output power of the selected generator.
Parameters:
<Level>
<numeric value>
*RST:
-20 dBm
Example:
SOUR:EXT:POW -30dBm
Sets the generator level to -30 dBm
Manual operation:
See
SOURce:EXTernal[:STATe]
<State>
This command activates or deactivates the connected external generator.
Parameters:
<State>
ON | OFF
*RST:
OFF
Manual operation:
See
SOURce:POWer[:LEVel][:IMMediate]:OFFSet
<Offset>
This command defines a level offset for the external generator level. Thus, for exam-
ple, attenuators or amplifiers at the output of the external generator can be taken into
account for the setting.
Parameters:
<Offset>
Range:
-200 dB to +200 dB
*RST:
0dB
Example:
SOUR:POW:OFFS -10dB
Sets the level offset of the external generator to – 20 dBm.
Usage:
SCPI confirmed
Manual operation:
See
Configuring the Measurement