Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
251
User Manual 1173.9240.02 ─ 17
Parameters:
<Source>
IMMediate
Free Run
EXTernal
Trigger signal from the TRIGGER INPUT connector.
EXT2
Trigger signal from the TRIGGER INPUT/OUTPUT connector.
Note: Connector must be configured for "Input".
EXT3
Trigger signal from the TRIGGER 3 INPUT/ OUTPUT connector.
Note: Connector must be configured for "Input".
RFPower
First intermediate frequency
Not available for input from the Digital Baseband Interface
(R&S
FSW-B17) or the Analog Baseband Interface (R&S
FSW-
B71).
IFPower
Second intermediate frequency
Not available for input from the Digital Baseband Interface
(R&S
FSW-B17). For input from the Analog Baseband Interface
(R&S
FSW-B71), this parameter is interpreted as
BBPower
for
compatibility reasons.
IQPower
Magnitude of sampled I/Q data
For applications that process I/Q data, such as the I/Q Analyzer
or optional applications.
Not available for input from the Digital Baseband Interface
(R&S
FSW-B17) or the Analog Baseband Interface (R&S
FSW-
B71).
TIME
Time interval
BBPower
Baseband power (for digital input via the Digital Baseband Inter-
face R&S
FSW-B17)
Baseband power (for digital input via the Digital Baseband Inter-
face R&S
FSW-B17 or the Analog Baseband interface
R&S
FSW-B71)
PSEN
External power sensor
AF
AF power signal
FM
FM power signal
AM
corresponds to the RF power signal
AMRelative
Configuring the Measurement