Configuration
R&S
®
FSW-K7
85
User Manual 1173.9240.02 ─ 17
For details on supported trigger levels, see the data sheet.
Remote command:
Hysteresis ← Using the power sensor as an external trigger
Defines the distance in dB to the trigger level that the trigger source must exceed
before a trigger event occurs. Setting a hysteresis avoids unwanted trigger events
caused by noise oscillation around the trigger level.
Remote command:
Trigger Holdoff ← Using the power sensor as an external trigger
Defines the minimum time (in seconds) that must pass between two trigger events.
Trigger events that occur during the holdoff time are ignored.
Remote command:
Drop-Out Time ← Using the power sensor as an external trigger
Defines the time the input signal must stay below the trigger level before triggering
again.
Slope ← Using the power sensor as an external trigger
Defines whether triggering occurs when the signal rises to the trigger level or falls
down to it.
Remote command:
5.4.3 Amplitude
The amplitude is configured in the "Amplitude" tab of the "Input/Frontend" dialog box.
Amplitude settings are identical to the base unit.
For background information on amplitude settings see the R&S
FSW User Manual.
► To display this dialog box, do one of the following:
● Select the "Input/Frontend" button in the Analog Demodulation "Overview" and
switch to the "Amplitude" tab.
● Select the AMPT key and then the "Amplitude Config" softkey.
Amplitude settings for input from the Analog Baseband interface (R&S
FSW-B71) are
chapter 5.4.3.2, "Amplitude Settings for Analog Baseband Input"
5.4.3.1
Amplitude Settings for RF Input
Amplitude settings can be configured via the AMPT key or in the "Amplitude" dialog
box.
Input and Frontend Settings