Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
205
User Manual 1173.9240.02 ─ 17
Manual operation:
See
INPut:DIQ:RANGe:COUPling
<State>
If enabled, the reference level for digital input is adjusted to the full scale level automat-
ically if the full scale level changes.
This command is only available if the optional Digital Baseband Interface (R&S
FSW-
B17) is installed.
Parameters:
<State>
ON | OFF
*RST:
OFF
Manual operation:
See
"Adjust Reference Level to Full Scale Level"
INPut:DIQ:RANGe[:UPPer]
<Level>
Defines or queries the "Full Scale Level", i.e. the level that corresponds to an I/Q sam-
ple with the magnitude "1".
This command is only available if the optional Digital Baseband Interface (R&S
FSW-
B17) is installed.
Parameters:
<Level>
<numeric value>
Range:
1 μV to 7.071 V
*RST:
1 V
Manual operation:
See
INPut:DIQ:RANGe[:UPPer]:UNIT
<Unit>
Defines the unit of the full scale level (see
bility of units depends on the measurement application you are using.
This command is only available if the optional Digital Baseband Interface (R&S
FSW-
B17) is installed.
Parameters:
<Level>
VOLT | DBM | DBPW | WATT | DBMV | DBUV | DBUA | AMPere
*RST:
Volt
Manual operation:
See
INPut:DIQ:SRATe
<SampleRate>
This command specifies or queries the sample rate of the input signal from the Digital
Baseband Interface (R&S
Configuring the Measurement