Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
296
User Manual 1173.9240.02 ─ 17
Parameters:
<THD>
Total harmonic distortion of the demodulated signal in dB.
Usage:
Query only
[SENSe:]ADEMod:FM:OFFSet?
<ResultType>
This command calculates the FM carrier offset from the currently available measure-
ment data set.
If averaging has been activated before acquiring the data set (using
on page 272, the averaged FM offset over sev-
eral measurements can also be obtained by setting <ResultType> = AVERage.
The offset thus determined differs from the one calculated by the
since, for determination of the frequency deviation, the modulation is removed by
means of low pass filtering, producing results that are different from those obtained by
averaging.
Query parameters:
<ResultType>
IMMediate | AVERage
IMMediate
The current measurement results are used to calculate the FM
offset
AVERage
The measurement results that were averaged over the given
number of measurements are used to calculate the FM offset
If no average measurement was active during the last measure-
ment sequence only the
[SENSe:]ADEMod:FM:OFFSet?
IMMediate
command will return a correct result (data to calcu-
late the offset are taken from the last measured data set).
[SENSe:]ADEMod:FM:OFFSet? AVERage
will cause a query
error in this case.
Example:
ADEM:SET 8MHz,32000,EXT,POS,-500,30
Sets up demodulator parameters to execute 30 measurements
ADEM:FM AVER,OFF,OFF
Selects FM results to perform averaging
INIT; WAI
Starts measurement and waits for sync
ADEM:FM:OFFS? IMM
Reads FM offset of last measurement of the sequence of 30
ADEM:FM:OFFS? AVER
Reads FM offset averaged over 30 measurements
Usage:
Query only
Retrieving Results