Configuration
R&S
®
FSW-K7
70
User Manual 1173.9240.02 ─ 17
The allowed range is from 100 Hz to 10 GHz.
Remote command:
Full Scale Level
The "Full Scale Level" defines the level and unit that should correspond to an I/Q sam-
ple with the magnitude "1".
If "Auto" is selected, the level is automatically set to the value provided by the connec-
ted device.
Remote command:
Adjust Reference Level to Full Scale Level
If enabled, the reference level is adjusted to the full scale level automatically if any
change occurs.
Remote command:
Connected Instrument
Displays the status of the Digital Baseband Interface connection.
If an instrument is connected, the following information is displayed:
●
Name and serial number of the instrument connected to the Digital Baseband Inter-
face
●
Used port
●
Sample rate of the data currently being transferred via the Digital Baseband Inter-
face
●
Level and unit that corresponds to an I/Q sample with the magnitude "1" (
), if provided by connected instrument
Remote command:
DigIConf
Starts the optional R&S DigIConf application. This softkey is available in the In-/Output
menu, but only if the optional software is installed.
Note that R&S DigIConf requires a USB connection (not LAN!) from the
R&S
FSW to the R&S EX-IQ-BOX in addition to the Digital Baseband Interface
(R&S
FSW-B17) connection. R&S DigIConf version 2.20.360.86 Build 170 or
higher is required.
To return to the R&S
FSW application, press any key on the front panel. The
R&S
FSW application is displayed with the "Input/Output" menu, regardless of which
key was pressed.
For details on the R&S DigIConf application, see the "R&S®EX-IQ-BOX Digital Inter-
face Module R&S®DigIConf Software Operating Manual".
Input and Frontend Settings