Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
316
User Manual 1173.9240.02 ─ 17
Parameters:
<SortMode>
X
Sorts the peaks according to increasing position on the x-axis.
Y
Sorts the peaks according to decreasing position on the y-axis.
*RST:
X
Example:
CALC:MARK:FUNC:FPE:SORT Y
Sets the sort mode to decreasing y values
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:STAT
<State>
This command turns a peak search on and off.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:MARK:FUNC:FPE:STAT ON
Activates marker peak search
Manual operation:
See
CALCulate:MARKer:FUNCtion:FPEeaks:X?
This command queries the position of the peaks on the x-axis.
The order depends on the sort order that has been set with
.
Return values:
<PeakPosition>
Position of the peaks on the x-axis. The unit depends on the
measurement.
Usage:
Query only
CALCulate:MARKer:FUNCtion:FPEeaks:Y?
This command queries the position of the peaks on the y-axis.
The order depends on the sort order that has been set with
.
Return values:
<PeakPosition>
Position of the peaks on the y-axis. The unit depends on the
measurement.
Usage:
Query only