Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
335
User Manual 1173.9240.02 ─ 17
1 = origin of coordinate system (x1 = 0, y1 = 0)
2 = end point of system (x2 = 100, y2= 100)
3 = zoom area (e.g. x1 = 60, y1 = 30, x2 = 80, y2 = 75)
Parameters:
<x1>,<y1>,
<x2>,<y2>
Diagram coordinates in % of the complete diagram that define
the zoom area.
The lower left corner is the origin of coordinate system. The
upper right corner is the end point of the system.
Range:
0 to 100
Default unit: PCT
Manual operation:
See
DISPlay[:WINDow<n>]:ZOOM:STATe
<State>
This command turns the zoom on and off.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
DISP:ZOOM ON
Activates the zoom mode.
Manual operation:
See
See
See
"Deactivating Zoom (Selection mode)"
11.8.3.2
Using the Multiple Zoom
DISPlay[:WINDow<n>]:ZOOM:MULTiple<zoom>:AREA
.....................................................335
DISPlay[:WINDow<n>]:ZOOM:MULTiple<zoom>:STATe
................................................... 336
DISPlay[:WINDow<n>]:ZOOM:MULTiple<zoom>:AREA
<x1>,<y1>,<x2>,<y2>
This command defines the zoom area for a multiple zoom.
To define a zoom area, you first have to turn the zoom on.
Analyzing Results