
Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
226
User Manual 1173.9240.02 ─ 17
Parameters:
<MeasType>
THRough
"TRANsmission" mode: calibration with direct connection
between external generator and device input
"REFLection" mode: calibration with short circuit at the input
OPEN
only allowed in "REFLection" mode: calibration with open input
Example:
INIT:CONT OFF
Selects single sweep operation
CORR:METH TRAN
Selects a transmission measurement.
CORR:COLL THR;*WAI
Starts the measurement of reference data using direct connec-
tion between generator and device input and waits for the sweep
end.
Usage:
Setting only
SCPI confirmed
Manual operation:
See
See
[SENSe:]CORRection:METHod
This command selects the type of measurement to be performed with the external gen-
erator.
This command is only available if external generator control is active (see
Parameters:
REFLection
Selects reflection measurements.
TRANsmission
Selects transmission measurements.
*RST:
TRANsmission
Example:
CORR:METH TRAN
Sets the type of measurement to "transmission".
Manual operation:
See
See
See
[SENSe:]CORRection:RECall
This command restores the measurement configuration used for calibration.
This command is only available if external generator control is active (see
Configuring the Measurement