Configuration
R&S
®
FSW-K7
113
User Manual 1173.9240.02 ─ 17
and f
max
=DBW/2
f
max
and span
min
are specified in the data sheet.
Remote command:
Demodulation Bandwidth
Defines the demodulation bandwidth of the measurement. The demodulation band-
width determines the sampling rate with which the input signal is captured and ana-
lyzed.
For recommendations on finding the correct demodulation bandwidth see
For details on the relation between demodulation bandwidth and sampling rate refer to
chapter 4.3, "Sample Rate, Measurement Time and Trigger Offset"
Remote command:
RF Full Span
Sets the span (around the center frequency) of the RF data to be evaluated to the
demodulation bandwidth.
Remote command:
5.8.3 AF Filter
The AF filter reduces the evaluated bandwidth of the demodulated signal and can
define a weighting function. It is configured in the "AF Filter" tab of the "Demodulation
Settings" dialog box.
AF filters are only available for AF time domain evaluations.
A maximum of two filters out of high pass, low pass or deemphasis filters can be active
at the same time if analog demodulation output is active (see
► To display this dialog box, do one of the following:
● Select the "Demod Settings" button in the Analog Demodulation "Overview"
and select the "AF Filter" tab.
● Select the MEAS CONFIG key and then the "AF Filter Config" softkey.
Demodulation