Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
243
User Manual 1173.9240.02 ─ 17
Example:
BAND:DEM 1MHz
Sets demodulation bandwidth to 1 MHz
Manual operation:
See
[SENSe:]BANDwidth|BWIDth:DEMod:TYPE
<FilterType>
This command defines the type of demodulation filter to be used.
This command is identical to
SENS:ADEM:BAND:DEM:TYPE
:
Parameters:
<FilterType>
FLAT
Standard flat demodulation filter
GAUSs
Gaussian filter for optimized settling behaviour
*RST:
FLAT
Manual operation:
See
[SENSe:]BANDwidth|BWIDth[:RESolution]
<Bandwidth>
This command defines the resolution bandwidth and decouples the resolution band-
width from the span.
For statistics measurements, this command defines the
demodulation
bandwidth.
For measurements on I/Q data in the frequency domain, the maximum RBW is 1
MHz.
Parameters:
<Bandwidth>
refer to data sheet
*RST:
RBW: AUTO is set to ON; DBW: 3MHz
Example:
BAND 1 MHz
Sets the resolution bandwidth to 1 MHz
Usage:
SCPI confirmed
Manual operation:
See
[SENSe:]SWEep:COUNt
<SweepCount>
This command defines the number of sweeps that the application uses to average
traces.
In case of continuous sweeps, the application calculates the moving average over the
average count.
In case of single sweeps, the application stops the measurement and calculates the
average after the average count has been reached.
Configuring the Measurement