Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
238
User Manual 1173.9240.02 ─ 17
Parameters:
<Gain>
15 dB | 30 dB
The availability of preamplification levels depends on the
R&S
FSW model.
R&S
FSW8/13: 15dB and 30 dB
R&S
FSW26 or higher: 30 dB
All other values are rounded to the nearest of these two.
*RST:
OFF
Example:
INP:GAIN:VAL 30
Switches on 30 dB preamplification.
Usage:
SCPI confirmed
Manual operation:
See
11.4.5.4
Scaling the Y-Axis
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]
........................................................................238
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:AUTO ONCE
....................................................238
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:MODE
............................................................. 238
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:RPOSition
........................................................239
DISPlay[:WINDow<n>]:TRACe:Y:SPACing
.......................................................................239
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]
<Range>
This command defines the display range of the y-axis.
Parameters:
<Range>
If the y-axis shows the power, the unit is dB with a range from
10
dB to 200
dB.
If the y-axis shows the frequency, the unit is Hz with a variable
range.
*RST:
100 dB (frequency domain), 500 kHz (time domain)
Example:
DISP:TRAC:Y 110dB
Usage:
SCPI confirmed
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:AUTO ONCE
Automatic scaling of the y-axis is performed once, then switched off again.
Usage:
SCPI confirmed
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:MODE
<Mode>
This command selects the type of scaling of the y-axis.