![Rohde & Schwarz FSW Series User Manual Download Page 60](http://html.mh-extra.com/html/rohde-and-schwarz/fsw-series/fsw-series_user-manual_1477844060.webp)
Configuration
R&S
®
FSW-K7
60
User Manual 1173.9240.02 ─ 17
Handover Freq.
Defines the frequency at which the mixer switches from one range to the next (if two
different ranges are selected). The handover frequency can be selected freely within
the overlapping frequency range.
Remote command:
Band
Defines the waveguide band or user-defined band to be used by the mixer.
The start and stop frequencies of the selected band are displayed in the "RF Start" and
"RF Stop" fields.
For a definition of the frequency range for the pre-defined bands, see
The mixer settings for the user-defined band can be selected freely. The frequency
range for the user-defined band is defined via the harmonics configuration (see
Remote command:
RF Overrange
If enabled, the frequency range is not restricted by the band limits ("RF Start" and "RF
Stop"). In this case, the full LO range of the selected harmonics is used.
Remote command:
Preset Band
Restores the presettings for the selected band.
Note:
changes to the band and mixer settings are maintained even after using the
PRESET function. This function allows you to restore the original band settings.
Remote command:
Mixer Type
The R&S
FSW option B21 supports the following external mixer types:
"2 Port"
LO and IF data use the same port
"3 Port"
LO and IF data use separate ports
Remote command:
Mixer Settings (Harmonics Configuration)
The harmonics configuration determines the frequency range for user-defined bands
(see
Input and Frontend Settings