Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
237
User Manual 1173.9240.02 ─ 17
INPut:EATT:STATe
<State>
This command turns the electronic attenuator on and off.
This command is only available with option R&S
FSW-B25.
It is not available if R&S
FSW-B17 is active.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
INP:EATT:STAT ON
Switches the electronic attenuator into the signal path.
Manual operation:
See
"Using Electronic Attenuation (Option B25)"
11.4.5.3
Configuring a Preamplifier
INPut:GAIN:STATe
<State>
This command turns the preamplifier on and off.
The command requires option R&S
FSW-B24.
This function is not available for input from the Digital Baseband Interface (R&S
FSW-
B17).
For R&S
FSW 26 or higher models, the input signal is amplified by 30 dB if the pream-
plifier is activated.
For R&S
FSW 8 or 13 models, the preamplification is defined by
.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
INP:GAIN:STAT ON
Switches on 30 dB preamplification.
Usage:
SCPI confirmed
Manual operation:
See
INPut:GAIN[:VALue]
<Gain>
This command selects the preamplification level if the preamplifier is activated
(
INP:GAIN:STAT ON
, see
The command requires option R&S
FSW-B24.
Configuring the Measurement