Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
323
User Manual 1173.9240.02 ─ 17
CALCulate:LIMit<k>:CONTrol[:DATA]
<LimitLinePoints>
This command defines the horizontal definition points of a limit line.
Parameters:
<LimitLinePoints>
Variable number of x-axis values.
Note that the number of horizontal values has to be the same as
the number of vertical values set with
not, the R&S
FSW either adds missing values or ignores surplus
values.
The unit is Hz or s.
*RST:
-
Usage:
SCPI confirmed
Manual operation:
See
CALCulate:LIMit<k>:CONTrol:DOMain
<SpanSetting>
This command selects the domain of the limit line.
Parameters:
<SpanSetting>
FREQuency | TIME
*RST:
FREQuency
Manual operation:
See
CALCulate:LIMit<k>:CONTrol:MODE
<Mode>
This command selects the horizontal limit line scaling.
Parameters:
<Mode>
ABSolute
Limit line is defined by absolute physical values (Hz or s).
RELative
Limit line is defined by relative values related to the center fre-
quency (frequency domain) or the left diagram border (time
domain).
*RST:
ABSolute
CALCulate:LIMit<k>:CONTrol:OFFSet
<Offset>
This command defines an offset for a complete limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Analyzing Results