Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
339
User Manual 1173.9240.02 ─ 17
......................................................................................339
.................................................................................... 339
CALCulate:RTMS:WINDow<n>:IVAL?
..............................................................................339
................................................................................... 339
CALCulate:RTMS:ALINe:SHOW
This command defines whether or not the analysis line is displayed in all time-based
windows in all MSRT applications and the MSRT Master.
Note
: even if the analysis line display is off, the indication whether or not the currently
defined line position lies within the analysis interval of the active application remains in
the window title bars.
Parameters:
<State>
ON | OFF
*RST:
ON
Manual operation:
See
CALCulate:RTMS:ALINe[:VALue]
<Position>
This command defines the position of the analysis line for all time-based windows in all
MSRT applications and the MSRT Master.
Parameters:
<Position>
Position of the analysis line in seconds. The position must lie
within the measurement time (pret posttrigger) of the
MSRT measurement.
Default unit: s
Manual operation:
See
CALCulate:RTMS:WINDow<n>:IVAL?
This command queries the analysis interval for the window specified by the index <n>.
This command is only available in application measurement channels, not the MSRT
View or MSRT Master.
Return values:
<IntStart>
Start value of the analysis interval in seconds
Default unit: s
<IntStop>
Stop value of the analysis interval in seconds
Usage:
Query only
[SENSe:]RTMS:CAPTure:OFFSet
<Offset>
This setting is only available for applications in MSRT mode, not for the MSRT Master.
It has a similar effect as the trigger offset in other measurements.
Analyzing Results