Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
217
User Manual 1173.9240.02 ─ 17
Parameters:
<Unit>
DB | PCT
*RST:
DB
Example:
UNIT:PMET:POW:RAT DB
Manual operation:
See
Triggering with Power Sensors
[SENSe:]PMETer<p>:TRIGger:DTIMe
..............................................................................217
[SENSe:]PMETer<p>:TRIGger:HOLDoff
...........................................................................217
[SENSe:]PMETer<p>:TRIGger:HYSTeresis
...................................................................... 218
[SENSe:]PMETer<p>:TRIGger:LEVel
...............................................................................218
[SENSe:]PMETer<p>:TRIGger:SLOPe
.............................................................................218
[SENSe:]PMETer<p>:TRIGger[:STATe]
........................................................................... 219
[SENSe:]PMETer<p>:TRIGger:DTIMe
<Time>
This command defines the time period that the input signal has to stay below the IF
power trigger level before the measurement starts.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Time>
Range:
0 s to 1 s
Increment: 100 ns
*RST:
100 µs
Example:
PMET2:TRIG:DTIMe 0.001
[SENSe:]PMETer<p>:TRIGger:HOLDoff
<Holdoff>
This command defines the trigger holdoff for external power triggers.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Holdoff>
Time period that has to pass between the trigger event and the
start of the measurement, in case another trigger event occurs.
Range:
0 s to 1 s
Increment: 100 ns
*RST:
0 s
Example:
PMET2:TRIG:HOLD 0.1
Sets the holdoff time of the trigger to 100 ms
Manual operation:
See
Configuring the Measurement