Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
341
User Manual 1173.9240.02 ─ 17
Manual operation:
See
MMEMory:STORe:IQ:STATe
1, <FileName>
This command writes the captured I/Q data to a file.
The file extension is *.iq.tar. By default, the contents of the file are in 32-bit floating
point format.
Secure User Mode
In secure user mode, settings that are to be stored on the instrument are stored to vol-
atile memory, which is restricted to 256
MB. Thus, a "Memory full" error may occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S
FSW User Manual.
Parameters:
1
<FileName>
String containing the path and name of the target file.
Example:
MMEM:STOR:IQ:STAT 1, 'C:
\R_S\Instr\user\data.iq.tar'
Stores the captured I/Q data to the specified file.
Manual operation:
See
11.10 Commands for Compatibility
The following commands are provided for compatibility to other signal analyzers only.
For new remote control programs use the specified alternative commands.
CALCulate<n>:FEED
<Evaluation>
This command selects the evaluation method of the measured data that is to be dis-
played in the specified window.
Note that this command is maintained for compatibility reasons only. Use the
LAYout
commands for new remote control programs (see
chapter 11.6.2, "Working with Win-
Parameters:
<Evaluation>
Type of evaluation you want to display.
See the table below for available parameter values.
Commands for Compatibility