Measurement Time (AQT)
Defines how long data is acquired for demodulatation. For details on the measurement
time see
chapter 4.3, "Sample Rate, Measurement Time and Trigger Offset"
Remote command:
Capture Offset
This setting is only available for applications in
MSRA or MSRT operating mode
. It
has a similar effect as the trigger offset in other measurements: it defines the time off-
set between the capture buffer start and the start of the extracted application data.
In MSRA mode, the offset must be a positive value, as the capture buffer starts at the
trigger time = 0.
In MSRT mode, the offset may be negative if a pretrigger time is defined.
Remote command:
Resolution Bandwidth
Defines the resolution bandwidth for data acquisition. The available range is specified
in the data sheet.
Remote command:
5.6.2 Sweep Settings
The sweep settings define how often data from the input signal is acquired and then
demodulated. They are configured via the SWEEP key or in the "Sweep" tab of the
"Data Acquisition" dialog box.
► To display this dialog box, do one of the following:
● Select the "Data Acquisition" button in the Analog Demodulation "Overview"
and switch to the "Sweep" tab.
● Select the SWEEP key and then the "Sweep Config" softkey.
Data Acquisition