Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
207
User Manual 1173.9240.02 ─ 17
[SENSe:]PROBe<p>:ID:SRNumber?
Queries the serial number of the probe.
Suffix:
<p>
.
1 | 2 | 3
Selects the connector:
1 = Baseband Input I
2 = Baseband Input Q
3 = RF (currently not supported; use "1" with RF Input Connec-
tor setting "Baseband Input I")
Return values:
<SerialNo>
Serial number in a string.
Usage:
Query only
[SENSe:]PROBe<p>:SETup:MODE
<Mode>
Select the action that is started with the micro button on the probe head.
See also:
Suffix:
<p>
.
1 | 2 | 3
Selects the connector:
1 = Baseband Input I
2 = Baseband Input Q
3 = RF (currently not supported; use "1" with RF Input Connec-
tor setting "Baseband Input I")
Parameters:
<Mode>
RSINgle
Run single: starts one data acquisition.
NOACtion
Nothing is started on pressing the micro button.
*RST:
RSINgle
Manual operation:
See
[SENSe:]PROBe<p>:SETup:NAME?
Queries the name of the probe.
Suffix:
<p>
.
1 | 2 | 3
Selects the connector:
1 = Baseband Input I
2 = Baseband Input Q
3 = RF (currently not supported; use "1" with RF Input Connec-
tor setting "Baseband Input I")
Configuring the Measurement