Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
208
User Manual 1173.9240.02 ─ 17
Return values:
<Name>
Name string
Usage:
Query only
[SENSe:]PROBe<p>:SETup:STATe?
Queries if the probe at the specified connector is active (detected) or not active (not
detected). To switch the probe on, i.e. activate input from the connector, use
INP:SEL:AIQ
(see
Suffix:
<p>
.
1 | 2 | 3
Selects the connector:
1 = Baseband Input I
2 = Baseband Input Q
3 = RF (currently not supported; use "1" with RF Input Connec-
tor setting "Baseband Input I")
Return values:
<State>
DETected | NDETected
*RST:
NDETected
Usage:
Query only
[SENSe:]PROBe<p>:SETup:TYPE?
Queries the type of the probe.
Suffix:
<p>
.
1 | 2 | 3
Selects the connector:
1 = Baseband Input I
2 = Baseband Input Q
3 = RF (currently not supported; use "1" with RF Input Connec-
tor setting "Baseband Input I")
Return values:
<Type>
String containing one of the following values:
–
None (no probe detected)
–
active differential
–
active single-ended
Usage:
Query only
11.4.2.6
Working with Power Sensors
The following commands describe how to work with power sensors.
.................................................................................. 209
Configuring Power Sensor Measurements
........................................................... 210
............................................................................. 217
Configuring the Measurement