Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
278
User Manual 1173.9240.02 ─ 17
Example:
ABOR;:INIT:IMM
Aborts the current measurement and immediately starts a new
one.
Example:
ABOR;*WAI
INIT:IMM
Aborts the current measurement and starts a new one once
abortion has been completed.
Usage:
SCPI confirmed
INITiate:CONMeas
This command restarts a (single) measurement that has been stopped (using
INIT:CONT OFF
) or finished in single sweep mode.
The measurement is restarted at the beginning, not where the previous measurement
was stopped.
, this command does not reset traces in
maxhold, minhold or average mode. Therefore it can be used to continue measure-
ments using maxhold or averaging functions.
Manual operation:
See
INITiate:CONTinuous
<State>
This command controls the sweep mode.
Note that in single sweep mode, you can synchronize to the end of the measurement
with *OPC, *OPC? or *WAI. In continuous sweep mode, synchronization to the end of
the measurement is not possible. Thus, it is not recommended that you use continuous
sweep mode in remote control, as results like trace data or markers are only valid after
a single sweep end synchronization.
For details on synchronization see the "Remote Basics" chapter in the R&S
FSW User
Manual.
If the sweep mode is changed for a measurement channel while the Sequencer is
active (see
on page 280) the mode is only con-
sidered the next time the measurement in that channel is activated by the Sequencer.
Parameters:
<State>
ON | OFF | 0 | 1
ON | 1
Continuous sweep
OFF | 0
Single sweep
*RST:
1
Capturing Data and Performing Sweeps