![Rohde & Schwarz FSW Series User Manual Download Page 225](http://html.mh-extra.com/html/rohde-and-schwarz/fsw-series/fsw-series_user-manual_1477844225.webp)
Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
225
User Manual 1173.9240.02 ─ 17
Parameters:
<Address>
TCP/IP address between 0.0.0.0 and 0.255.255.255
*RST:
0.0.0.0
Example:
SYST:COMM:TCP:RDEV:GEN:ADDR 130.094.122.195
Manual operation:
See
"GPIB Address / TCP/IP Address"
Source Calibration
The following commands are required to activate the calibration functions of the exter-
nal tracking generator. However, they are only available if external generator control is
active (see
Remote commands exclusive to source calibration:
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:RVALue
........................................................... 225
[SENSe:]CORRection:COLLect[:ACQuire]
........................................................................225
....................................................................................... 226
..........................................................................................226
........................................................................................ 227
[SENSe:]CORRection:TRANsducer:GENerator
.................................................................227
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:RVALue
<Value>
The command defines the power value assigned to the reference position in the grid.
For external generator calibration measurements (requires External Generator Control
option R&S
FSW-B10), this command defines the power offset value assigned to the
reference position.
Parameters:
<Value>
*RST:
0 dBm, coupled to reference level
Example:
DISP:TRAC:Y:RVAL -20dBm
Sets the power value assigned to the reference position to -20
dBm
Manual operation:
See
[SENSe:]CORRection:COLLect[:ACQuire]
<MeasType>
This command initiates a reference measurement (calibration). The reference mea-
surement is the basis for the measurement normalization. The result depends on
whether a reflection measurement or transmission measurement is performed (see
To obtain a correct reference measurement, a complete sweep with synchronization to
the end of the sweep must have been carried out. This is only possible in the single
sweep mode.
This command is only available if external generator control is active (see
Configuring the Measurement