5.5.1 Trigger Source Settings
The trigger source settings are configured in the "Trigger Source" tab of the "Trigger"
dialog box.
..................................................................................... 95
.............................................................................................96
............................................................................ 97
................................................................................................. 98
Trigger Source
In the Analog Demodulation application, the next measurement can be triggered if the
selected input signal exceeds the threshold specified using the "Trigger Level" setting
(see
on page 98). Thus, a periodic signal modulated onto the carrier
frequency can be displayed. It is recommended that the measurement time covers at
least five periods of the audio signal.
Remote command:
Free Run ← Trigger Source
No trigger source is considered. Data acquisition is started manually or automatically
and continues until stopped explicitely.
Remote command:
TRIG:SOUR IMM
, see
External Trigger 1/2/3 ← Trigger Source
Data acquisition starts when the TTL signal fed into the specified input connector (on
the front or rear panel) meets or exceeds the specified trigger level.
(See
Note:
The "External Trigger 1" softkey automatically selects the trigger signal from the
TRIGGER INPUT connector on the front panel.
For details see the "Instrument Tour" chapter in the R&S
FSW Getting Started manual.
"External Trigger 1"
Trigger signal from the TRIGGER 1 INPUT connector on the front
panel.