
Analysis
R&S
®
FSW-K7
146
User Manual 1173.9240.02 ─ 17
........................................................................................146
........................................................................................147
............................................................................................... 147
Peak List State
Activates/deactivates the marker peak list. If activated, the peak list is displayed and
the peaks are indicated in the trace display.
For each listed peak the frequency/time ("X-value") and level ("Y-value") values are
given.
Remote command:
Sort Mode
Defines whether the peak list is sorted according to the x-values or y-values. In either
case the values are sorted in ascending order.
Remote command:
Maximum Number of Peaks
Defines the maximum number of peaks to be determined and displayed.
Remote command:
Peak Excursion
Defines the minimum level value by which a signal must rise or fall so that it will be
identified as a maximum or a minimum by the search functions.
Marker Peak List Configuration