Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
317
User Manual 1173.9240.02 ─ 17
MMEMory:STORe:LIST
<FileName>
This command exports the SEM and spurious emission list evaluation to a file.
The file format is *.dat.
Secure User Mode
In secure user mode, settings that are to be stored on the instrument are stored to vol-
atile memory, which is restricted to 256
MB. Thus, a "Memory full" error may occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S
FSW User Manual.
Parameters:
<FileName>
String containing the path and name of the target file.
Example:
MMEM:STOR:LIST 'test'
Stores the current list evaluation results in the
test.dat
file.
MMEMory:STORe:PEAK
<FileName>
This command exports the marker peak list to a file.
The file format is *.dat.
Secure User Mode
In secure user mode, settings that are to be stored on the instrument are stored to vol-
atile memory, which is restricted to 256
MB. Thus, a "Memory full" error may occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S
FSW User Manual.
Parameters:
<FileName>
String containing the path and name of the target file.
Example:
MMEM:STOR:PEAK 'test'
Saves the current marker peak list in the file
test.dat
.
Usage:
Event
Manual operation:
See
n dB Down Marker
The following commands control the n dB down markers.