![Rohde & Schwarz FSW Series User Manual Download Page 259](http://html.mh-extra.com/html/rohde-and-schwarz/fsw-series/fsw-series_user-manual_1477844259.webp)
Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
259
User Manual 1173.9240.02 ─ 17
Manual operation:
See
[SENSe:]ADEMod:AF:SPAN
<Span>
This command sets the span (around the center frequency) for AF spectrum result dis-
play.
The span is limited to DBW/2 (see
Parameters:
<Span>
*RST:
9 MHz
Example:
ADEM:AF:SPAN 200 kHz
Sets the AF span to 200 kHz
Manual operation:
See
[SENSe:]ADEMod:AF:SPAN:FULL
This command sets the maximum span for AF spectrum result display.
The maximum span corresponds to DBW/2 (see
Example:
ADEM:BAND 5 MHz
Sets the demodulation bandwidth to 5 MHz
ADEM:AF:SPAN:FULL
Sets the AF span to 2.5 MHz
Manual operation:
See
[SENSe:]ADEMod:AF:STARt
<Frequency>
This command sets the start frequency for AF spectrum result display.
Parameters:
<Frequency>
*RST:
0 MHz
Example:
ADEM:AF:STAR 0 kHz
Sets the AF start frequency to 0 kHz
ADEM:AF:STOP 500 kHz
Sets the AF stop frequency to 500 kHz
Manual operation:
See
[SENSe:]ADEMod:AF:STOP
<Frequency>
This command sets the stop frequency for AF spectrum result display.
Parameters:
<Frequency>
*RST:
9 MHz