![Rohde & Schwarz FSW Series User Manual Download Page 307](http://html.mh-extra.com/html/rohde-and-schwarz/fsw-series/fsw-series_user-manual_1477844307.webp)
Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
307
User Manual 1173.9240.02 ─ 17
Parameters:
<StepSize>
STANdard
the marker moves from one pixel to the next
POINts
the marker moves from one sweep point to the next
*RST:
POINts
Example:
CALC:MARK:X:SSIZ STAN
Sets the marker step size to one pixel.
Manual operation:
See
CALCulate:MARKer<m>:LINK
<DisplayType>
Links the specified marker in all displays of the specified type.
Parameters:
<DisplayType>
TIME | SPECtrum | BOTH | NONE
TIME
Links the markers in all time domain diagrams
SPECtrum
Links the markers in all AF Spectrum displays
BOTH
Links the markers both in the time domain diagrams and in the
AF Spectrum displays
NONE
Markers are not linked.
*RST:
NONE
Manual operation:
See
See
DISPlay:MTABle
<DisplayMode>
This command turns the marker table on and off.
Parameters:
<DisplayMode>
ON
Turns the marker table on.
OFF
Turns the marker table off.
AUTO
Turns the marker table on if 3 or more markers are active.
*RST:
AUTO
Example:
DISP:MTAB ON
Activates the marker table.
Manual operation:
See
Analyzing Results