![Rohde & Schwarz FSW Series User Manual Download Page 247](http://html.mh-extra.com/html/rohde-and-schwarz/fsw-series/fsw-series_user-manual_1477844247.webp)
Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
247
User Manual 1173.9240.02 ─ 17
Example:
TRIG:SOUR IFP
Sets the IF power trigger source.
TRIG:IFP:HYST 10DB
Sets the hysteresis limit value.
Manual operation:
See
TRIGger[:SEQuence]:LEVel:BBPower
<Level>
This command sets the level of the baseband power trigger.
This command is available for the
Digital Baseband Interface (R&S
FSW-B17)
and
the
Analog Baseband Interface (R&S
FSW-B71)
.
Parameters:
<Level>
Range:
-50 dBm to +20 dBm
*RST:
-20 dBm
Example:
TRIG:LEV:BB -30DBM
TRIGger[:SEQuence]:LEVel[:EXTernal<port>]
<TriggerLevel>
This command defines the level the external signal must exceed to cause a trigger
event.
Note that the variable INPUT/OUTPUT connectors (ports 2+3) must be set for use as
input using the
Suffix:
<port>
.
Selects the trigger port.
1 = trigger port 1 (TRIGGER INPUT connector on front panel)
2 = trigger port 2 (TRIGGER INPUT/OUTPUT connector on front
panel)
3 = trigger port 3 (TRIGGER3 INPUT/OUTPUT connector on
rear panel)
Parameters:
<TriggerLevel>
Range:
0.5 V to 3.5 V
*RST:
1.4 V
Example:
TRIG:LEV 2V
Manual operation:
See
TRIGger[:SEQuence]:LEVel:IFPower
<TriggerLevel>
This command defines the power level at the third intermediate frequency that must be
exceeded to cause a trigger event. Note that any RF attenuation or preamplification is
considered when the trigger level is analyzed. If defined, a reference level offset is also
considered.
For compatibility reasons, this command is also available for the "baseband power"
trigger source when using the Analog Baseband Interface (R&S
FSW-B71).
Configuring the Measurement