Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
320
User Manual 1173.9240.02 ─ 17
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:TIME?
This command queries the position of the n dB down markers on the x-axis when
measuring in the time domain.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also
Return values:
<TimeX1>
absolute position in time of the n dB marker to the left of the ref-
erence marker in seconds
<TimeX2>
absolute position in time of the n dB marker to the right of the
reference marker in seconds
Example:
INIT:CONT OFF
Switches to single sweep mode
CALC:MARK:FUNC:NDBD ON
Switches on the n dB down function.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:NDBD:TIME?
Outputs the time values of the temporary markers.
Usage:
Query only
Manual operation:
See
Phase Noise Measurement Marker
The following commands control the phase noise measurement marker function.
Useful commands for phase noise markers described elsewhere
●
●
●
Remote commands exclusive to phase noise markers
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:AUTO
.............................................. 320
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:RESult?
........................................... 321
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise[:STATe]
........................................... 321
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:AUTO
<State>
This command turns an automatic peak search for the fixed reference marker at the
end of a sweep on and off.
Analyzing Results