Analysis
R&S
®
FSW-K7
140
User Manual 1173.9240.02 ─ 17
●
Press the MKR key, then select the "Marker Config" softkey. Then select the hori-
zontal "Search Settings" tab.
●
In the "Overview", select "Analysis", and switch to the vertical "Marker Config" tab.
Then select the horizontal "Search Settings" tab.
........................................................................................ 140
Search Mode for Next Peak
Selects the search mode for the next peak search.
"Left"
Determines the next maximum/minimum to the left of the current
peak.
"Absolute"
Determines the next maximum/minimum to either side of the current
peak.
"Right"
Determines the next maximum/minimum to the right of the current
peak.
Remote command:
Peak Excursion
Defines the minimum level value by which a signal must rise or fall so that it will be
identified as a maximum or a minimum by the search functions.
Entries from 0
dB to 80
dB are allowed; the resolution is 0.1
dB. The default setting for
the peak excursion is 6
dB.
Remote command:
6.4.2 Positioning Functions
The following functions set the currently selected marker to the result of a peak search
or set other characteristic values to the current marker value. These functions are
available as softkeys in the "Marker To" menu, which is displayed when you press the
MKR -> key, or in the "Marker" menu.
Marker Search Settings and Positioning Functions